Extrel

MAX300-AIR ENVIRONMENTAL MASS SPECTROMETER

The MAX300-AIR is an industrial gas analyzer optimized for environmental and ambient air monitoring and regulatory compliance. Quadrupole mass spectrometry is extremely sensitive and very fast, allowing for the rapid detection and quantitation of a wide variety of toxic industrial contaminants. The speed of the MAX300-AIR allows one analyzer to measure unlimited compounds and be automated to monitor 160+ sample points, spread throughout the manufacturing site.

Description

Description

The MAX300-AIR is an industrial gas analyzer optimized for environmental and ambient air monitoring and regulatory compliance. Quadrupole mass spectrometry is extremely sensitive and very fast, allowing for the rapid detection and quantitation of a wide variety of toxic industrial contaminants. The speed of the MAX300-AIR allows one analyzer to measure unlimited compounds and be automated to monitor 160+ sample points, spread throughout the manufacturing site.

MAX300-AIR Measurements

  • Toxic chemical release
  • Total sulfur
  • TWA Exposure Values
  • BTU (Energy content)
  • Explosive Limit Analysis
Features
  • Quantitative analysis of individual contaminants
  • ppt detection limits
  • Analysis time in seconds
  • Multiport sample systems for total site monitoring
  • High precision and accuracy for safety and regulatory reporting
  • Low maintenance and calibration required

System Specifications

Mass Range Options
  • 1-200, 300, 500 amu

Detectable Compounds

  • Any gas or vapor sample

Detection Range

  • 100% – 10 ppb standard, 10 ppt w/ membrane inlet*

Number of Sample Streams

  • 16, 31, 40, 80, 120, 160+

Analysis Rate

  • 0.4 Seconds per component

Number of Components

  • Unlimited
Number of Analysis Routines
  • Unlimited

Number of User Configurable Data Tags

  • Unlimited

Analysis Precision

  • ±0.25% Relative Standard Deviation**

Stability

  • ±0.5% Relative Standard Deviation**

Filaments

  • Two, one active and one spare with automatic switchover

Maintenance

  • Minimal
  • Manual or fully automated calibration and validation
19 mm High-transmission Quadrupole filter
*Matrix dependent. Documented on trace benzene in air**Based on the analysis of 1% argon


Product Notes

MAX300-AIR PRODUCT NOTE IPA100A 


Documentation


Publications

Title

Go to Top